Details

The DRE Citadel VS provides vessel sealing and bipolar electrosurgery capabilities. With this ESU, clinicians can create precise cuts using all kinds of modes: Vessel Sealing, Bipolar Cut & Coag, and Desiccate. The Citadel VS also includes smart 6SENSE™ technology, along with a range of features to ensure optimal workflow and patient safety. 

The DRE Citadel VS electrosurgical generator is available through DRE. Contact an DRE representative today for more information! 

Features 

  • Smart ENDO-CUT mode for endoscopic applications.
  • 5” LCD touch-screen
  • Five bipolar modes (two cut, three coag)
  • Two vessel sealing modes
  • Dual paddle footswitch to toggle between monopolar and bipolar modes
  • Cool seal technology included to create a more translucent seal
  • 30 user-programmable settings to cover surgical procedures with more flexibility
  • 6SENSE™ technology gives advanced feedback which senses changes in voltage, current, power, tissue density, and leakage RF current
  • Optional time out function to stop the HF output on unintentional prolonged activation

Specifications

Vessel Sealer

  • Fine Seal: 150 W at 100 Ω, CF 1.5
  • Rapid Seal: 150 W at 100 Ω, CF 1.5

Bipolar

  • Cut - Macro: 100 W at 100 Ω, CF 1.5
  • Cut - BiCut: 100 W at 100 Ω, CF 1.5
  • Coag - Micro (Precise): 100 W at 100 Ω, CF 1.5
  • Coag - Standard: 100 W at 100 Ω, CF 1.5
  • Coag - Force: 100 W at 100 Ω, CF 1.5

Desiccate

  • Desiccate: 150 W at 100 Ω, CF 1.5

Dimensions

  • Length: 38 cm (15”)
  • Width: 30.5 cm (12”)
  • Height: 11.5 cm (4.5”)
  • Weight: < 5 kg (11 lb.)

Power – 230 V

  • Input mains voltage, full regulation range: 210-264 Vac
  • Input mains voltage, operating range: 180-264 Vac
  • Mains current (maximum): Idle: 0.2 A | Bipolar: 2 A | Sealer: 2 A
  • Maximum VA at nominal line voltage: Idle: 52 VA | Bipolar: 528VA | Sealer: 528 VA
  • Mains line frequency range (nominal): 50 to 60 Hz
  • Fusing: 6A

Power – 110 V

  • Input mains voltage, full regulation range: 104-132 Vac
  • Input mains voltage, operating range: 85-135 Vac
  • Mains current (maximum): Idle: 0.4 A | Bipolar: 4 A | Sealer: 4 A
  • Maximum VA at nominal line voltage: Idle: 52 VA | Bipolar: 528VA | Sealer: 528 VA
  • Mains line frequency range (nominal): 50 to 60 Hz
  • Fusing: 10A

Open Circuit P-P Voltage (Volts)

  • Bipolar Cut – Macro: 700
  • Bipolar Cut – BiCut: 700
  • Bipolar Coag – Micro: 380
  • Bipolar Coag – Standard: 560
  • Bipolar Coag – Force: 700
  • Sealer – Rapid Seal: 560
  • Sealer – Fine Seal: 560
  • Desiccate: 560